5310C Lab TOC Analyzer
Measures TOC Samples - All Day, Every Day

To learn more about the 5310C Oxidation & Measuring System through an animated tour, click on button below.....

Download Brochure

The Sievers 5310 C Laboratory TOC Analyzer, a member of the Sievers 900 Series, was developed specifically for drinking water applications. The 5310 C has an analytical range of 30 parts-per-trillion to 50 parts-per-million and utilizes SM 5310 C for USEPA compliance monitoring of raw and finished drinking water. The 5310 C provides unsurpassed ease of use, industry-best analytical performance, and a low cost of ownership for water quality analysis.

The 5310 C offers superior productivity for laboratory TOC measurements. Automated calibration and verification procedures, 4-minute analysis time, and automated reagent adjustment combine for an unequalled level of efficiency, flexibility, and ease of use. The 5310 C can be used with the 900 Autosampler, which features convenient random access capability and up to 120 sample positions.

The 5310 C Laboratory was engineered for ease of use and cost-effective operation. It needs no external reagents or gas supplies and requires only a few hours of preventive maintenance each year. Twelve-month calibration stability frees analysts to perform other critical tasks. A large, color touch-screen display provides an intuitive menu for easy setup of instrument parameters. A small 19.2-cm (7.6-in) wide profile efficiently uses valuable bench space.

The Sievers 5310 C Analyzer uses the UV/persulfate oxidation method and patented Sievers Membrane Conductometric Detection technology, which delivers unmatched accuracy and precision across the Analyzer’s wide operating range

Operating Specifications*
Range 4 ppb to 50 ppm
Precision < 1% RSD
Accuracy ± 2% or ± 0.5 ppb, whichever is greater
Display Readout 3 significant digits
Sample Type Autosampler or discrete grab sample
Calibration Typically stable for 12 months
Analysis Time 4 minutes
Sample Temperature 1–95 °C (34–203 °F)
Ambient Temperature 10–40 °C (50–104 °F)
Instrument Sample
Flow Rate
0.5 mL/min
Instrument Specifications
Outputs Serial RS-232 output (2); USB port (1); parallel printer port (1)
Display Quarter-VGA, color touch-sensitive LCD
Power Universal Power Supply: 100–240 ± 10% VAC, 50 W, 50/60 Hz
Dimensions H: 48.3 cm (19.0 in); W: 19.2 cm (7.6 in); D: 48.0 cm (18.9 in)
Weight 14.3 kg (31.5 lb)
Safety Certifications ETL, CE
Consumables
UV Lamp 6 months
Acid Reagent As needed, typically 6-12 months depending on usage
Oxidizer Reagent As needed, typically 3-month stability; available in 150- or 300-mL cartridge
Pump Tubing 12 months
Resin Bed 12 months

* Stated analytical performance is achievable under controlled laboratory conditions that minimize operator and standards errors.