5310C Lab TOC Analyzer
Measures TOC Samples - All Day, Every Day
To learn more about the 5310C Oxidation & Measuring System through an animated tour, click on button below.....
The Sievers 5310 C Laboratory TOC Analyzer, a member of the Sievers 900 Series, was developed specifically for drinking water applications. The 5310 C has an analytical range of 30 parts-per-trillion to 50 parts-per-million and utilizes SM 5310 C for USEPA compliance monitoring of raw and finished drinking water. The 5310 C provides unsurpassed ease of use, industry-best analytical performance, and a low cost of ownership for water quality analysis.
The 5310 C offers superior productivity for laboratory TOC measurements. Automated calibration and verification procedures, 4-minute analysis time, and automated reagent adjustment combine for an unequalled level of efficiency, flexibility, and ease of use. The 5310 C can be used with the 900 Autosampler, which features convenient random access capability and up to 120 sample positions.
The 5310 C Laboratory was engineered for ease of use and cost-effective operation. It needs no external reagents or gas supplies and requires only a few hours of preventive maintenance each year. Twelve-month calibration stability frees analysts to perform other critical tasks. A large, color touch-screen display provides an intuitive menu for easy setup of instrument parameters. A small 19.2-cm (7.6-in) wide profile efficiently uses valuable bench space.
The Sievers 5310 C Analyzer uses the UV/persulfate oxidation method and patented Sievers Membrane Conductometric Detection technology, which delivers unmatched accuracy and precision across the Analyzer’s wide operating range
| Operating Specifications* | |
|---|---|
| Range | 4 ppb to 50 ppm |
| Precision | < 1% RSD |
| Accuracy | ± 2% or ± 0.5 ppb, whichever is greater |
| Display Readout | 3 significant digits |
| Sample Type | Autosampler or discrete grab sample |
| Calibration | Typically stable for 12 months |
| Analysis Time | 4 minutes |
| Sample Temperature | 1–95 °C (34–203 °F) |
| Ambient Temperature | 10–40 °C (50–104 °F) |
|
Instrument Sample Flow Rate |
0.5 mL/min |
| Instrument Specifications | |
| Outputs | Serial RS-232 output (2); USB port (1); parallel printer port (1) |
| Display | Quarter-VGA, color touch-sensitive LCD |
| Power | Universal Power Supply: 100–240 ± 10% VAC, 50 W, 50/60 Hz |
| Dimensions | H: 48.3 cm (19.0 in); W: 19.2 cm (7.6 in); D: 48.0 cm (18.9 in) |
| Weight | 14.3 kg (31.5 lb) |
| Safety Certifications | ETL, CE |
| Consumables | |
| UV Lamp | 6 months |
| Acid Reagent | As needed, typically 6-12 months depending on usage |
| Oxidizer Reagent | As needed, typically 3-month stability; available in 150- or 300-mL cartridge |
| Pump Tubing | 12 months |
| Resin Bed | 12 months |
* Stated analytical performance is achievable under controlled laboratory conditions that minimize operator and standards errors.