5310C On-Line TOC Analyser

5310C_OL_270.jpgEasy operation, Low maintenance and Superior Accuracy  

The Sievers 5310 C On-Line Total Organic Carbon (TOC) Analyzer was designed specifically for the municipal drinking water market for continuous monitoring of raw, finished and plant process waters. It offers superior accuracy and precision across a dynamic operating range of 4 parts per billion (ppb) to 50 parts per million (ppm) TOC.

The 5310 C On-Line Analyzer was engineered for ease of use and cost-effective operation, just like its laboratory model, the Sievers 5310 C Laboratory. It simplifies monitoring and enhances productivity with automated procedures such as calibration, verification, reagent adjustment, and percent removal calculation. The low-maintenance 5310 C needs no external reagents or gas supplies, and offers 12-month calibration stability.

The Analyzer has a dust-resistant and spray-proof IP-45 rated enclosure for demanding water environments. It runs in continuous analysis mode, and can also accept discrete grab samples and standards at any time through the patented Sievers Integrated On-Line Sampling System (iOS System).

Operating Specifications*

Range 4 ppb to 50 ppm
Precision < 1% RSD
Accuracy ± 2% or ± 0.5 ppb, whichever is greater
Display Readout 3 significant digits
Sample Type On-line continuous or discrete grab sample
Calibration Typically stable for 12 months
Analysis Time 4 minutes
Sample Temperature** 1–95 °C (34–203 °F)
Ambient Temperature 10–40 °C (50–104 °F)
Sample Pressure** Up to 250 psi
Required Sample Line Flow Rate 50–300 mL/min (for on-line mode)
Instrument Sample Flow Rate 0.5 mL/min
Analyzer Specifications
Outputs 4-20mA output (1); Serial RS-232 output (2); USB port (1); parallel printer port (1)
Display Quarter-VGA, color touch-sensitive LCD
Power Universal Power Supply: 100–240 ± 10% VAC, 50 W, 50/60 Hz
Dimensions H: 62.4 cm (24.6 in); W: 45.2 cm (17.8 in); D: 26.4 cm (10.4 in)
Weight 16.9 kg (37.2 lb)
Safety Certifications ETL, CE
Enclosure Rating IP-45
Consumables
UV Lamp 6 months
Acid Reagent As needed, typically 3 months with an ICR or 6 months without an ICR
Oxidizer Reagent As needed, typically 3-month stability; available in 150- or 300-mL cartridge
Pump Tubing 12 months
Resin Bed 12 months
*Stated analytical performance is achievable under controlled laboratory conditions that minimize operator and standards errors
**The sample temperature and pressure should not simultaneously exceed 60 °C and 100 psi